These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: An isodose shift technique for obliquely incident electron beams. Author: Ulin K, Sternick ES. Journal: Med Phys; 1989; 16(6):905-10. PubMed ID: 2511399. Abstract: It is well known that when an electron beam is incident obliquely on the surface of a phantom, the depth dose curve measured normal to the surface is shifted toward the surface. Based on geometrical arguments alone, the depth of the nth isodose line for an electron beam incident at an angle theta should be equal to the product of cos theta and the depth of the nth isodose line at normal incidence. This method, however, ignores the effects of scatter and can lead to significant errors in isodose placement for beams at large angles of incidence. A semi-empirical functional relationship and a table of isodose shift factors have been developed with which one may easily calculate the depth of any isodose line for beams at incident angles of 0 degrees to 60 degrees. The isodose shift factors are tabulated in terms of beam energy (6-22 MeV) and isodose line (10%-90%) and are shown to be relatively independent of beam size and incident angle for angles less than 60 degrees. Extensive measurements have been made on a Varian Clinac 2500 linear accelerator with a parallel-plate chamber and polystyrene phantom. The dependence of the chamber response on beam angulation has been checked, and the scaling factor of the polystyrene phantom has been determined to be equal to 1.00.[Abstract] [Full Text] [Related] [New Search]