These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Comparative study of parathormone and vitamin D measurements by three automats: ADVIA Centaur XP® (Siemens), ISYS® (IDS) and Liaison® (Diasorin)]. Author: Gardien P, Moineau MP, Kerspern H, Bordron A, Carre JL. Journal: Ann Biol Clin (Paris); 2014; 72(4):443-52. PubMed ID: 25119802. Abstract: Vitamin D (Vit.D) and parathormone (PTH) measurements are usually prescribed for phosphocalcic metabolism assessment and, especially for Vit.D, more and more frequently for other pathologies. In order to step up to automated techniques for these analysis in our laboratory, we tested 3 devices: ADVIA Centaur XP(®) (Siemens), ISYS(®) (IDS) and Liaison(®) (Diasorin), which allow to simultaneously quantify Vit.D and PTH. The aim of this study was to evaluate the fidelity of these methods as well as study the correlation between them and the radioimmunological techniques previously used in our laboratory : « ELSA PTH » (Iba Cisbio International) and « 25-OH D » (IDS). The comparison of PTH analysis was performed on a population of chronic renal failure patients undergoing haemodialysis. According to our study, the 3 devices show acceptable analytical performances; anyway the measurements realized on the ISYS analyzer are the ones showing the best results in terms of fidelity, and the closest results to those obtained with the RIA reference techniques.[Abstract] [Full Text] [Related] [New Search]