These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Standard endovascular treatment of abdominal aortic aneurysms in patients with very short proximal necks using the Endurant stent graft.
    Author: Matsagkas M, Kouvelos G, Peroulis M, Avgos S, Arnaoutoglou E, Papa N, Papadopoulos G.
    Journal: J Vasc Surg; 2015 Jan; 61(1):9-15. PubMed ID: 25124361.
    Abstract:
    OBJECTIVE: This study evaluated and compared the midterm results of endovascular aortic aneurysm repair with the Endurant (Medtronic Inc, Santa Rosa, Calif) stent graft system in off-label use in patients with short (<10 mm) proximal aortic necks and in patients treated according to device-specific instructions for use. METHODS: This was a case-control (2:1) single-center retrospective analysis of prospectively collected data performed between September 2008 and December 2012. Analysis identified 19 elective patients with short (<10 mm) proximal necks and mild angulations (≤45°) treated with the Endurant stent graft and 38 patients matched for age, sex, and aneurysm diameter with proximal aortic necks ≥10 mm in length who met the instructions for use. End points included technical and clinical success and freedom from any secondary intervention, any type of endoleak, and aneurysm-related death. RESULTS: The short-neck group was a mean ± standard deviation age of 71.7 ± 8.9 years, 84% were men, and their mean infrarenal aortic neck length was 6.1 ± 1.2 mm. Mean suprarenal and infrarenal angles were 110° ± 10.4° and 170° ± 15.4°, respectively. Aortic neck diameters were similar between the groups (26.6 ± 3.8 vs 25.7 ± 3.7 mm; P = .36). Primary technical success was achieved in all cases. Off-label patients were more likely to require additional proximal cuff deployment to successfully obtain a seal (21% vs 3%; P = .04). The two patient groups were similar in rates of perioperative mortality, morbidity, and complications. Mean follow-up of 24 ± 12 months revealed no differences in clinical success, freedom from reintervention, and aneurysm-related death. No type I endoleaks were observed in either group during the follow-up period. CONCLUSIONS: The Endurant stent graft system applied off-label in patients with very short aneurysm necks (<10 mm) with mild angulation showed acceptable treatment results. These midterm results might suggest its use in carefully selected patients with very short neck anatomy. Long-term data are needed to verify the observed durability of the Endurant stent graft.
    [Abstract] [Full Text] [Related] [New Search]