These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Comparative Analysis of Various Aspects of Plateletpheresis on the Fenwal Amicus and Fresenius COM.TEC Cell Separator Instruments.
    Author: Philip J, Biswas AK, Chatterjee T, Mallhi RS.
    Journal: Lab Med; 2014; 45(4):315-23. PubMed ID: 25316663.
    Abstract:
    OBJECTIVE: To compare the Fenwal Amicus and the Fresenius COM.TEC apheresis instruments regarding donor peripheral blood parameters, operational variables of the instruments, and quality control parameters of the product obtained. METHODS: We performed 100 platelet collections from 100 voluntary donors using the 2 studied devices. We measured platelet count using an automated analyzer and analyzed the activation statuses using a flow cytometer. RESULTS: The median time needed to perform the procedures was significantly longer with the COM.TEC. However, the product we obtained using the Amicus instrument showed higher degrees of platelet-activation. All products we obtained with both instruments had white blood cell counts of less than 5 × 10(6) per bag. We observed no statistical difference regarding collection efficiency and collection rates between the devices. CONCLUSION: Both instruments collected platelets efficiently, with minimal donor discomfort. Compared with the COM.TEC instrument, the Amicus reached the platelet target yield more quickly; however, it displayed an increase in platelet activation.
    [Abstract] [Full Text] [Related] [New Search]