Search MEDLINE/PubMed
Title: Comparative Analysis of Various Aspects of Plateletpheresis on the Fenwal Amicus and Fresenius COM.TEC Cell Separator Instruments. Author: Philip J, Biswas AK, Chatterjee T, Mallhi RS. Journal: Lab Med; 2014; 45(4):315-23. PubMed ID: 25316663. Abstract: OBJECTIVE: To compare the Fenwal Amicus and the Fresenius COM.TEC apheresis instruments regarding donor peripheral blood parameters, operational variables of the instruments, and quality control parameters of the product obtained. METHODS: We performed 100 platelet collections from 100 voluntary donors using the 2 studied devices. We measured platelet count using an automated analyzer and analyzed the activation statuses using a flow cytometer. RESULTS: The median time needed to perform the procedures was significantly longer with the COM.TEC. However, the product we obtained using the Amicus instrument showed higher degrees of platelet-activation. All products we obtained with both instruments had white blood cell counts of less than 5 × 10(6) per bag. We observed no statistical difference regarding collection efficiency and collection rates between the devices. CONCLUSION: Both instruments collected platelets efficiently, with minimal donor discomfort. Compared with the COM.TEC instrument, the Amicus reached the platelet target yield more quickly; however, it displayed an increase in platelet activation.[Abstract] [Full Text] [Related] [New Search]