These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Transporting neonates with hypoxic-ischemic encephalopathy utilizing active hypothermia. Author: Weiss MD, Tang A, Young L, Irwin L, Brophy C, Larsen V, Howard J, Miller C, Douglas-Escobar M. Journal: J Neonatal Perinatal Med; 2014 Jan 01; 7(3):173-8. PubMed ID: 25318630. Abstract: OBJECTIVE: The objective of this report was to evaluate a servo-controlled active hypothermia device used during the transport of neonates with HIE. STUDY DESIGN: Retrospective review of all cases of therapeutic hypothermia, both passive and active, using a servo-controlled device in a single regional referral neonatal intensive care unit from 2009-2013 RESULTS: An ambulance (43%), fixed wing aircraft (25%), or helicopter (32%) transported 28 neonates with active hypothermia. The servo-controlled device captured core temperatures in all 28 neonates, resulting in 2,985 minutes of data. All neonates attained a core temperature between 33-34 °C by 33 minutes of the transport. Once the neonates attained a core temperature, the average temperature for the remainder of the transport was 33 ± 0.2 °C. The neonates maintained the core temperature regardless of the type of transport vehicle. CONCLUSION: Servo-controlled hypothermia enables rapid attainment of targeted temperatures and maintains the temperature throughout the transport process in both ground and air transport.[Abstract] [Full Text] [Related] [New Search]