These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Development of glucose biosensor based on reconstitution of glucose oxidase onto polymeric redox mediator coated pencil graphite electrodes.
    Author: Dervisevic M, Cevik E, Senel M.
    Journal: Enzyme Microb Technol; 2015 Jan; 68():69-76. PubMed ID: 25435508.
    Abstract:
    In this study, a novel glucose biosensor was fabricated by reconstitutional immobilization of glucose oxidase (GOx) onto a poly(glycidyl methacrylate-co-vinylferrocene) (poly(GMA-co-VFc)) film coated pencil graphite electrode (PGE). The amperometric current response of poly(GMA-co-VFc)-GOx to glucose is linear in the concentration range between 1 and 16mM (correlation coefficient of 0.9998) with a detection limit of 2.7μM (S/N=3). Experimental parameters were studied in detail and optimized, including the pH and temperature governing the analytical performance of the biosensor. The stability and reusability of the biosensor as well as its kinetic parameters have also been studied.
    [Abstract] [Full Text] [Related] [New Search]