These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Real-time fluorescence Loop-Mediated Isothermal Amplification (LAMP) for rapid and reliable diagnosis of pulmonary tuberculosis. Author: Cao D, Hu L, Lin M, Li M, Ye Z, Sun H, Huang J, Yang H, Tian J. Journal: J Microbiol Methods; 2015 Feb; 109():74-8. PubMed ID: 25533219. Abstract: A reliable, simple and rapid diagnostic method that can be helpful in pulmonary tuberculosis diagnosis is urgently needed. Loop-mediated Isothermal Amplification (LAMP) allows DNA to be amplified rapidly at a constant temperature. In this study, real-time fluorescence LAMP was evaluated to rapidly detect Mycobacterium tuberculosis in sputum and was compared to the performance of real-time fluorescence quantitative PCR (Q-PCR). All the standard MTB strains were successfully detected and limit of detection (LOD) was 10(2)CFU/mL by real-time fluorescence LAMP within 20min. In light of MTB in sputum, the real-time fluorescence LAMP method yielded a sensitivity of 98.0% and a specificity of 78.3%, compared to Q-PCR assay, which yielded a sensitivity of 96.0% and a specificity of 82.6% for PTB diagnosis. There was an excellent overall agreement between LAMP and Q-PCR for PTB (κ=0.315) and non-PTB (κ=0.862). Therefore, the real-time fluorescence LAMP assay is a rapid, sensitive, and specific method to detect pulmonary tuberculosis.[Abstract] [Full Text] [Related] [New Search]