Title: Arrhythmia detection in single-lead ECG by combining beat and rhythm-level information. Author: Pathangay V, Rath SP. Journal: Annu Int Conf IEEE Eng Med Biol Soc; 2014; 2014():3236-9. PubMed ID: 25570680. Abstract: In this paper, we propose a method for detecting arrhythmia in single-lead electro-cardiogram (ECG) signal. By applying a sequence of pre-processing steps (filtering, baseline correction), beat classification and rhythm identification, six different beat-types and four abnormal rhythms are detected. Beat classification uses fast Fourier transform (FFT) as the feature and a support vector machine (SVM) classifier. Subsequently rhythm identification uses a deterministic finite state machine to detect abnormal rhythms. We evaluate the performance of our technique on the MIT-BIH database, to obtain 97% beat classification accuracy and perfect rhythm identification result.[Abstract] [Full Text] [Related] [New Search]