These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A comparison of fibreoptic-guided tracheal intubation through the Ambu ® Aura-i ™, the intubating laryngeal mask airway and the i-gel ™: a manikin study. Author: de Lloyd LJ, Subash F, Wilkes AR, Hodzovic I. Journal: Anaesthesia; 2015 May; 70(5):591-7. PubMed ID: 25631299. Abstract: We compared the Aura-i(™) , intubating laryngeal mask airway and i-gel(™) as conduits for fibreoptic-guided tracheal intubation in a manikin. Thirty anaesthetists each performed two tracheal intubations through each device, a total of 180 intubations. The median (IQR [range]) time to complete the first intubation was 40 (31-50 [15-162]) s, 37 (34-48 [25-75]) s and 28 (22-35 [14-59]) s for the Aura-i, intubating laryngeal mask airway and i-gel, respectively. Tracheal intubation through the i-gel was the quickest (p < 0.01). Resistance to railroading of the tracheal tube over the fibrescope was significantly greater through the Aura-i compared with the intubating laryngeal mask airway and the i-gel (p = 0.001). There were no failures to intubate through the intubating laryngeal mask airway or the i-gel but six intubation attempts through the Aura-i were unsuccessful, in five owing to a railroading failure and in one owing to accidental oesophageal intubation. We conclude that the Aura-i does not perform as well as the intubating laryngeal mask airway or the i-gel as an adjunct for performing fibreoptic-guided tracheal intubation.[Abstract] [Full Text] [Related] [New Search]