These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Functional determination of C1 esterase inhibitor in fractions derived from plasma.
    Author: Benny AG, Brimer CM.
    Journal: Haematologia (Budap); 1989; 22(3):189-93. PubMed ID: 2583599.
    Abstract:
    A traditional esterolytic assay, a commercial chromogenic assay and an in-house chromogenic assay were compared for suitability in estimating C 1 esterase inhibitor (C 1 INH) activity in plasma and fractions obtained during preparation of a clinical concentrate of C 1 INH. The esterolytic and both of the chromogenic assays showed similar C 1 INH activity in unfractionated plasma samples. However, the esterolytic assay was significantly influenced by the buffer pH and concentration in fractionated plasma samples, and this often resulted in a significant overestimation of C 1 INH in some fractions. Similarly, the commercial C 1 INH assay kit did not always reliably estimate C 1 INH activity in fractions derived from plasma. A modified in-house functional assay was found to be more reliable in measuring C 1 INH in fractionated plasma samples. There was a good correlation between this activity assay and immunological assay estimates of C 1 INH in fractions derived from plasma.
    [Abstract] [Full Text] [Related] [New Search]