Title: [The analytical quality and comparability value of four glucometers from different manufacturers]. Author: Horovskaia LA, Lobachevskaia TV, Chernichuk OV. Journal: Klin Lab Diagn; 2015 Jan; 60(1):60-3. PubMed ID: 25874308. Abstract: The modern laboratory medicine provides observation of requirements to glucometers in line with standard DIN EN ISO 15197-2:2013 for self-control under treatment of patients with diabetes mellitus and registration of biologic types data. The study compared functioning of diagnostic devices in point-of-care of four manufacturers. The samples of whole blood taken from two patients were analyzed in three lots of test-strips with concentration of glucose 4.5 and 11.8 mmol/l using glucometers Accu-Chek Active, One Touch Select, Sattelite Express and Contour TS and reference analyzer Olympus AU 400. The results of measurements were analyzed using technique of separated test of patient. The results demonstrated that precision indicators of measurement of blood glucose on glucometers Accu-Chek Active and One Touch Select were within the scope of quality criteria DIN EN ISO 15197-2:2013. The results of testing of glucometer Sattelite Express exceeded the given criteria under measurement of normoglycemic level of concentration and for Contour TS in all range of measured values.[Abstract] [Full Text] [Related] [New Search]