These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [Control of the ultrafiltration rate as the main component of a new balancing system in continuous hemofiltration].
    Author: Stein B, Maucher H, Wiedeck H, Born B, Ahnefeld FW.
    Journal: Anaesthesist; 1989 Oct; 38(10):536-8. PubMed ID: 2589627.
    Abstract:
    In 27 patients undergoing continuous pump-driven hemofiltration during intensive care, a new balancing device was used to control transmembrane pressure difference, and thus ultrafiltration rate, by means of an elastic reservoir and an infusion pump. The simultaneous infusion of substitution fluid by a second pump working independently of the former provided exactly and automatically the intended fluid balance. In only 0.7% of the total hemofiltration time (5519 h) could the preset ultrafiltration rate not be achieved. The simple components and installation of this device recommend its application in every intensive care unit that uses continuous arteriovenous or continuous pump-driven hemofiltration, relieving the staff and reducing unwanted and unnoticed deviations of intravascular volume and fluid balance.
    [Abstract] [Full Text] [Related] [New Search]