These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Measurement of endogenous lithium levels in serum and urine by electrothermal atomic absorption spectrometry: a method with potential clinical applications. Author: Miller NL, Durr JA, Alfrey AC. Journal: Anal Biochem; 1989 Nov 01; 182(2):245-9. PubMed ID: 2610339. Abstract: A highly sensitive flameless atomic absorption method has been adapted for the determination of endogenous trace lithium levels in serum and urine. With ammonium nitrate as the only matrix modifier, serum levels of Li as low as 0.03 mumol/liter are measured accurately and there is no requirement for standard additions. The need for background correction during analysis was clearly established, and tungsten and Zeeman-effect background corrections were compared. The tungsten correction offered superior sensitivity and linearity of standards. Recoveries in urine and serum average 94.8 +/- 7.7 and 95.3 +/- 6.1% (+/- SD), respectively. The endogenous serum Li levels were 0.16 +/- 0.08 mumol/liter for normal subjects dwelling in the Denver metropolitan area. The mean 24-h excretion rate was 5.24 +/- 1.4 mumol/day. The mean fractional excretion of endogenous Li (clearance Li/clearance creatinine) was 23.2 +/- 3.0%, a value similar to values published for exogenously administered Li and measured by conventional methods.[Abstract] [Full Text] [Related] [New Search]