These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Guidelines for the automated evaluation of Elispot assays. Author: Janetzki S, Price L, Schroeder H, Britten CM, Welters MJ, Hoos A. Journal: Nat Protoc; 2015 Jul; 10(7):1098-115. PubMed ID: 26110715. Abstract: The presented protocol for Elispot plate evaluation summarizes how to implement the recommendations developed following the establishment of a large-scale international Elispot plate-reading panel and subsequent multistep consensus-finding process. The panel involved >100 scientists from various immunological backgrounds. The protocol includes the description and justification of steps for setting reading parameters to obtain accurate, reliable and precise automated analysis results of Elispot plates. Further, necessary adjustments for out-of-specification situations are described and examples are provided. The plate analysis, including parameter adjustments, auditing of results and necessary annotations, should be achievable within a time range of 10-30 min per plate. Adoption of these guidelines should enable a further reduction in assay variability and an increase in the reliability and comparability of results obtained by Elispot. These guidelines conclude the ongoing harmonization efforts for the enzymatic Elispot assay.[Abstract] [Full Text] [Related] [New Search]