These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: New fabrication technique for highly sensitive qPlus sensor with well-defined spring constant.
    Author: Labidi H, Kupsta M, Huff T, Salomons M, Vick D, Taucer M, Pitters J, Wolkow RA.
    Journal: Ultramicroscopy; 2015 Nov; 158():33-7. PubMed ID: 26117434.
    Abstract:
    A new technique for the fabrication of highly sensitive qPlus sensor for atomic force microscopy (AFM) is described. The focused ion beam was used to cut then weld onto a bare quartz tuning fork a sharp micro-tip from an electrochemically etched tungsten wire. The resulting qPlus sensor exhibits high resonance frequency and quality factor allowing increased force gradient sensitivity. Its spring constant can be determined precisely which allows accurate quantitative AFM measurements. The sensor is shown to be very stable and could undergo usual UHV tip cleaning including e-beam and field evaporation as well as in situ STM tip treatment. Preliminary results with STM and AFM atomic resolution imaging at 4.5 K of the silicon Si(111)-7×7 surface are presented.
    [Abstract] [Full Text] [Related] [New Search]