These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: In-hospital bedside blood glucose monitoring: the importance of a quality control program. Author: Davis SG, White NH, Stoverock LS, Santiago JV. Journal: J Pediatr Nurs; 1989 Oct; 4(5):353-6. PubMed ID: 2614651. Abstract: This study examined the adequacy of a quality control program to assure accuracy of blood glucose monitoring performed at the bedside by nurses in a general pediatric hospital. A standard quality control program during which all nurses received inservice training and demonstrated proficiency resulted in accuracies (percentage of values within 15% of the laboratory) of only 68.6% and 69.2% over two successive 3-month periods. A standardized refresher course did not improve accuracy. Expansion of the program to include identification and reinstruction of individual nurses who had more than 20% of their blood glucose measurements more than 10% off the laboratory value or any one value more than 40% off the laboratory during the previous quarter resulted in improved accuracies (83%, 78%, and 91%) over three successive 3-month periods. These findings indicate that a systemic quality control program including individual monitoring and remedial retraining is required to improve the long-term accuracy of bedside blood glucose monitoring by nurses in a general pediatric hospital.[Abstract] [Full Text] [Related] [New Search]