These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: EDEL: ENEA DOSEMETER FOR EYE LENS.
    Author: Ferrari P, Mariotti F, Campani L.
    Journal: Radiat Prot Dosimetry; 2016 Sep; 170(1-4):145-9. PubMed ID: 26170436.
    Abstract:
    Since the publication of International Commission on Radiological Protection statement in 2011 on tissue reaction, eye lens radiation protection played an important role in exposed personnel dosimetry. For this reason, the Italian National Agency for New Technologies, Energy and Sustainable Economic Development (ENEA) Individual Monitoring Service decided to study a prototype to fulfil specific requests (e.g. for survey in interventional department and intercomparisons). On the basis of such preliminary investigation, a new eye lens dosemeter was developed. The new dosemeter, named EDEL (ENEA Dosemeter for Eye Lens), was characterised in terms of Hp(3), the operational quantity related to eye lens monitoring. The investigation was performed experimentally and optimised using the Monte Carlo MCNP6 code. The new prototype was thought to fulfil two main requests: the reliability of the dosimetric data and the portability of the dosemeter itself. The new dosemeter will soon be supplied to the collaborating hospitals for workplace test measurements.
    [Abstract] [Full Text] [Related] [New Search]