    The main aim of this study was to analyze how screws affect the strain concentration induced on the mandibular condyle during implantation, screwing, and drilling, as well as after condylar loading. A clean cadaveric mandible was analyzed experimentally in the intact state and was then implanted with a Biomet/Lorenz Microfixation temporomandibular joint (TMJ) implant with seven bicortical self-tapping screws. The external surface of the mandible was instrumented with three strain gauges. A load of 500N on the TMJ was applied to the condyle before and after implantation. The results showed a strain concentration of -1500μɛ near the screws due to their implantation on the external surface of the mandible. The drilling process induced up to 80μɛ near the hole. The strain concentration did not change when there were more than six screws. Loading on the TMJ before and after implantation presented only a 10% difference in maximum principal strain. This study demonstrates the importance of the strain concentration induced by the screws. The process of implanting screws shows the importance of lateral surface preparation for a good fit in the condyle. Strain distribution after implantation and loading of the Biomet implant was found to be similar to that in the intact condyle.
    [Abstract] [Full Text] [Related] [New Search]