These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Subcutaneous Implantable Cardioverter Defibrillator Lead Failure due to Twiddler Syndrome.
    Author: Kooiman KM, Brouwer TF, Van Halm VP, Knops RE.
    Journal: Pacing Clin Electrophysiol; 2015 Nov; 38(11):1369-71. PubMed ID: 26306701.
    Abstract:
    We present a case of Twiddler syndrome in a patient with a subcutaneous implantable cardioverter defibrillator (S-ICD). The patient presented herself to the outpatient clinic with pain in the left chest. Chest x-ray confirmed Twiddler syndrome and ICD read-out revealed lead failure resulting in absent heart rhythm sensing in one vector. The lead and pulse generator were extracted and a new S-ICD system was reimplanted submuscular underneath the serratus anterior muscle to prevent reoccurrence. Lead investigation revealed an insulation defect caused by excessive mechanical stress.
    [Abstract] [Full Text] [Related] [New Search]