These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Mechanisms and Patterns of Intravascular Ultrasound In-Stent Restenosis Among Bare Metal Stents and First- and Second-Generation Drug-Eluting Stents. Author: Goto K, Zhao Z, Matsumura M, Dohi T, Kobayashi N, Kirtane AJ, Rabbani LE, Collins MB, Parikh MA, Kodali SK, Leon MB, Moses JW, Mintz GS, Maehara A. Journal: Am J Cardiol; 2015 Nov 01; 116(9):1351-7. PubMed ID: 26341188. Abstract: The most common causes of in-stent restenosis (ISR) are intimal hyperplasia and stent under expansion. The purpose of this study was to use intravascular ultrasound (IVUS) to compare the ISR mechanisms of bare metal stents (BMS), first-generation drug-eluting stents (DES), and second-generation DES. There were 298 ISR lesions including 52 BMS, 73 sirolimus-eluting stents, 52 paclitaxel-eluting stents, 16 zotarolimus-eluting stents, and 105 everolimus-eluting stent. Mean patient age was 66.6 ± 1.1 years, 74.2% were men, and 48.3% had diabetes mellitus. BMS restenosis presented later (70.0 ± 66.7 months) with more intimal hyperplasia compared with DES (BMS 58.6 ± 15.5%, first-generation DES 52.6 ± 20.9%, second-generation DES 48.2 ± 22.2%, p = 0.02). Although reference lumen areas were similar in BMS and first- and second-generation DES, restenotic DES were longer (BMS 21.8 ± 13.5 mm, first-generation DES 29.4 ± 16.1 mm, second-generation DES 32.1 ± 18.7 mm, p = 0.003), and stent areas were smaller (BMS 7.2 ± 2.4 mm(2), first-generation DES 6.1 ± 2.1 mm(2), second-generation DES 5.7 ± 2.0 mm(2), p <0.001). Stent fracture was seen only in DES (first-generation DES 7 [5.0%], second-generation DES 8 [7.4%], p = 0.13). In conclusion, restenotic first- and second-generation DES were characterized by less neointimal hyperplasia, smaller stent areas, longer stent lengths, and more stent fractures than restenotic BMS.[Abstract] [Full Text] [Related] [New Search]