PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Cardiac Troponin I carryover by very high patient samples still causes false-positive results on the Beckman Coulter AccuTnI + 3. Author: Wilgen U, Pretorius CJ, Gould MJ, Ungerer JP. Journal: Ann Clin Biochem; 2016 Jan; 53(Pt 1):177-9. PubMed ID: 26345181. Abstract: BACKGROUND: False-positive cardiac troponin I results as a result of carryover have previously been reported on the Beckman Coulter AccuTnI assay. We sought to determine if the carryover problem had been resolved with the new AccuTnI + 3 assay. METHODS: Carryover experiments were performed in parallel on the Beckman Coulter Access2 analyser using the legacy AccuTnI and new AccuTnI + 3 assays. The same negative patient pool sample was analysed before and after a single analysis of an extremely elevated patient sample. RESULTS: Analysis of a single extremely high sample caused elevations above the 99th percentile cut-off, and thus false-positive cardiac troponin I results on both assays. Both assays demonstrated carryover and subsequent further elevations in negative pool results the following day. CONCLUSIONS: Our study replicates our previously published findings of carryover and reagent pack contamination on the AccuTnI assay. Despite improvements on the new AccuTnI + 3 assay, carryover and reagent pack contamination are still present.[Abstract] [Full Text] [Related] [New Search]