These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Stabilizing the capsular bag and expanding the pupil with a pupil expansion device. Author: Zarei-Ghanavati S, Bagherian H. Journal: J Cataract Refract Surg; 2015 Sep; 41(9):1801-3. PubMed ID: 26471052. Abstract: UNLABELLED: We describe a surgical technique for management of zonular instability and pupillary miosis using a pupil expansion device (Malyugin ring) in patients who need cataract surgery. In this technique, the pupil expansion device is first used in the usual fashion. After a capsulorhexis has been created, lateral scrolls of the pupil expansion device are released and repositioned to entrap both the anterior capsulorhexis and the pupil margin at the against-the-wound meridian (3 o'clock and 9 o'clock position) to fixate the unstable capsular bag to the iris. Phacoemulsification is performed, and the ring is removed after intraocular lens implantation. With this method, capsular stabilization and pupil expansion are achieved simultaneously with the same instrument. FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.[Abstract] [Full Text] [Related] [New Search]