These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Comparison of 3 latest-generation glucose micro-reflectometers]. Author: Sehling-Werle KM, Werle E, Hageloch W, Huber KH, Pollert R, Weicker H. Journal: Schweiz Med Wochenschr; 1989 Apr 05; 119(13-14):440-5. PubMed ID: 2655073. Abstract: Three glucose reflectance meters (Reflolux II = Accu-Chek II, Glucometer II and Hypocount GA) were tested for precision and accuracy when used by medical personnel of a diabetic outpatient department and for self-monitoring at home. In addition, the visual readability of the appropriate reagent strip was checked. All three systems were sufficiently valid for reflectometric reading, while visual evaluation showed a higher deviation. The precision of Accu-Chek II under optimal conditions in the outpatient department was comparable to the precision of laboratory examinations (CV 3.3%). When used by patients themselves, Accu-Chek II and Glucometer II were sufficiently precise (CV 4.8% [corrected] and 5.3% respectively). These devices are recommended for blood glucose self-monitoring.[Abstract] [Full Text] [Related] [New Search]