These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: THE IMPACT OF X-RAY UNIT TYPE USED FOR ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURES ON PATIENT DOSES. Author: Tsapaki V, Paraskeva KD, Tsalafoutas IA, Paspatis G, Scotiniotis H, Georgopoulos P, Finou P, Karagianis JA. Journal: Radiat Prot Dosimetry; 2016 Dec; 171(4):503-508. PubMed ID: 26582174. Abstract: To investigate whether the X-ray unit type used for interventional endoscopic retrograde cholangiopancreatography (ERCP) procedures may affect patient radiation doses. A total of 471 ERCP procedures performed in 4 hospitals with 4 types of X-ray units were studied. Kerma-area product (KAP), fluoroscopy time (T) and total number of radiographs acquired (F) were recorded. KAP, T and F values exhibited a great variation, ranging from 0.1 to 130.2 Gy cm2 (mean 16 Gy cm2), 0.13 to 33.7 min (mean 5.4 min) and 0 to 26 radiographs (mean 3.5), respectively. The respective mean values for the four types of X-ray units that were investigated were as follows: KAP: 17.4, 12.5, 5.6 and 36.3 Gy cm2, T: 4.7, 5.2, 3.8 and 11.5 min and F: 1.7, 7.4, 1.9 and 4.6 radiographs. The type of the X-ray unit seems to significantly affect patient radiation dose, with the C-arm delivering the lowest and the angiography unit the highest patient doses.[Abstract] [Full Text] [Related] [New Search]