These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: A real-time autoregressive spectrum analyzer for Doppler ultrasound signals.
    Author: Schlindwein FS, Evans DH.
    Journal: Ultrasound Med Biol; 1989; 15(3):263-72. PubMed ID: 2662553.
    Abstract:
    A system based on a digital signal processor and a microcomputer has been programmed to estimate the maximum entropy autoregressive (AR) power spectrum of ultrasonic Doppler shift signals and display the results in the form of a sonogram in real-time on a computer screen. The system, which is based on a TMS 320C25 digital signal processor chip, calculates spectra with 128 frequency components from 64 samples of the Doppler signal. The samples are collected at a programmable rate of up to 40.96 kHz, and the computation of each spectrum takes typically 3.2 ms. The feasibility of on-line AR spectral estimation makes this type of analysis an attractive alternative to the more conventional fast Fourier transform approach to the analysis of Doppler ultrasound signals.
    [Abstract] [Full Text] [Related] [New Search]