These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Comparison of the transcutaneous bilirubinometers BiliCare and Minolta JM-103 in late preterm and term neonates.
    Author: Pratesi S, Boni L, Tofani L, Berti E, Sollai S, Dani C.
    Journal: J Matern Fetal Neonatal Med; 2016 Sep; 29(18):3014-8. PubMed ID: 26632858.
    Abstract:
    OBJECTIVES: To assess the agreement of transcutaneous bilirubin (TcB) measurement with the Bilicare™ System in comparison to TcB measured with JM-103™ and total serum bilirubin (TSB). METHODS: Caucasian infants with gestational age ≥35 weeks with non-hemolytic jaundice received TcB measurement with both Bilicare™ and JM-103™ devices. TSB was also obtained in infants at risk of phototherapy. RESULTS: We studied 458 infants measuring TcB with Bilicare™ and JM-103™, correlating the results and with TSB. The mean difference ± 2SD between Bilicare™ and JM-103™ TcB was 2.02 ± 4.46 mg/dL and decreased from 2.88 ± 3.17 to 1.20 ± 4.55, and to -0.95 ± 4.58 mg/dL at mild, moderate and high TcB values, respectively. CONCLUSIONS: Bilicare™ and JM-103™ TcB measurements are well correlated, but Bilicare™ over-estimates TcB for mild and moderate values and under-estimates it for high values compared to JM-103™. This could increase the prescription of TSB measurements for less serious cases and decrease them in the most worrisome.
    [Abstract] [Full Text] [Related] [New Search]