These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A sandwich-type electrochemical immunosensor for carcinoembryonic antigen based on signal amplification strategy of optimized ferrocene functionalized Fe₃O₄@SiO₂ as labels. Author: Feng T, Qiao X, Wang H, Sun Z, Hong C. Journal: Biosens Bioelectron; 2016 May 15; 79():48-54. PubMed ID: 26686923. Abstract: A sandwich-type electrochemical immunosensor was developed for sensitive detection of carcinoembryonic antigen (CEA) by using ferroferric oxide@silica-amino groups (Fe3O4@SiO2-NH2) as carriers and gold nanoparticles-graphene oxide (GO-AuNPs) as platform. The Fe3O4@SiO2-NH2 surface was used as linked reagents for co-immobilization of ferrocenecarboxylic acid (Fc-COOH) and secondary anti-CEA (Ab2) to prepare the signal probe, and it also could hasten the decomposition of hydrogen peroxide (H2O2) to amplify signals. Differential pulse voltammetry (DPV) was successfully used to quantify CEA. Under the optimized conditions, the designed immunosensor shows an excellent analytical performance wide dynamic response range of CEA concentration from 0.001 ng mL(-1) to 80 ng mL(-1) with a relatively low detection limit of 0.0002 ng mL(-1) (S/N=3), and high specificity and good reproducibility. The proposed immunosensor was successfully used to determine CEA in spiked human serum samples.[Abstract] [Full Text] [Related] [New Search]