These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Reduction of scatter-induced image noise in cone beam computed tomography: effect of field of view size and position. Author: Pauwels R, Jacobs R, Bogaerts R, Bosmans H, Panmekiate S. Journal: Oral Surg Oral Med Oral Pathol Oral Radiol; 2016 Feb; 121(2):188-95. PubMed ID: 26792756. Abstract: OBJECTIVE: To measure the effect of field of view (FOV) size and position on scatter-induced image noise in cone beam computed tomography (CBCT). STUDY DESIGN: A polymethyl methacrylate (PMMA) phantom containing air and aluminum underwent CBCT scanning, using seven FOVs ranging between 4 × 4 cm and 14 × 10 cm, positioned both centrally and according to a dental scan. Signal difference to noise ratio (SDNR) was measured on two-dimensional (2-D) projection images. RESULTS: At a central position, SDNR decreased with increasing FOV size, ranging between 9.8 (14 × 10 cm) and 10.9 (4 × 4 cm). For dental FOV positions, SDNR values were between 6.3 (14 × 10 cm) and 9.5 (4 × 4 cm). To reach a constant SDNR, a dose reduction up to 76% was possible for small FOVs compared with the 14 × 10 cm FOV. CONCLUSIONS: The use of small FOVs and peripheral FOV positioning decreases scatter at the detector, resulting in a considerable potential for reduction of radiation dose to the patient.[Abstract] [Full Text] [Related] [New Search]