These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Pre-clinical Experience with a Multi-Chordal Patch for Mitral Valve Repair.
    Author: Chawla SK, Shi W, McIver BV, Vinten-Johansen J, Frater RW, Padala M.
    Journal: J Cardiovasc Transl Res; 2016 Apr; 9(2):127-34. PubMed ID: 26801477.
    Abstract:
    Surgical repair of flail mitral valve leaflets with neochordoplasty has good outcomes, but implementing it in anterior and bi-leaflet leaflet repair is challenging. Placing and sizing individual neochordae is time consuming and error prone, with persistent localized flail if performed incorrectly. In this study, we report our pre-clinical experience with a novel multi-chordal patch for mitral valve repair. The device was designed based on human cadaver hearts, and laser cut from expanded polytetrafluoroethylene. The prototypes were tested in: (stage 1) ex vivo hearts with leaflet flail (N = 6), (stage 2) acute swine induced with flail (N = 6), and (stage 3) two chronic swine survived to 23 and 120 days (N = 2). A2 and P2 prolapse were successfully repaired with coaptation length restored to 8.1 ± 2.2mm after posterior repair and to 10.2 ± 1.3mm after anterior repair in ex vivo hearts. In vivo, trace regurgitation was seen after repair with excellent patch durability, healing, and endothelialization at euthanasia. A new device for easier mitral repair is reported, with good early pre-clinical outcomes.
    [Abstract] [Full Text] [Related] [New Search]