These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Evaluation of time-resolved fluorometric immunoassay of CA 50]. Author: Hara K, Mitsuhashi F, Sakawaki T, Ohkura H. Journal: Rinsho Byori; 1989 Jul; 37(7):789-94. PubMed ID: 2691736. Abstract: Serum CA 50 was determined by a time resolved fluorometric immunoassay (TR-FIA) with CANAG CA-50 DELFIA kit. Evaluation of the assay system gave satisfactory results in its sensitivity, accuracy, reproducibility, dynamic range and easy handling. No prozone phenomenon was observed up to 347,000 U/ml. From a histogram of 134 normal sera, the cut off point was determined at 34 U/ml. CA 50 in 202 patients' sera was determined with this assay. Nineteen of 20 patients pancreatic cancer, 6 of 21 gastric cancer, 14 of 25 hepatoma gave positive values. In comparison with CA 19-9, higher values and higher rates of positive CA 50 were observed in benign and malignant liver diseases, suggesting its non-cancerous origin in the liver. A high correlation was observed between the level of CA 50 and CA 19-9 of 157 patients' sera. Serum CA 50 was completely correlated with CA 19-9 in the clinical course of patients with pancreatic cancer, but not in patients with hepatoma. Thus we conclude that the CANAG CA-50 DELFIA System is useful for the diagnosis and monitoring cancer patients but must be used with care because of its elevation in benign liver diseases.[Abstract] [Full Text] [Related] [New Search]