  • Title: A Novel Conductive Poly(3,4-ethylenedioxythiophene)-BSA Film for the Construction of a Durable HRP Biosensor Modified with NanoAu Particles.
    Author: Xu F, Ren S, Gu Y.
    Journal: Sensors (Basel); 2016 Mar 15; 16(3):. PubMed ID: 26999133.
    Abstract:
    In this study, we have investigated the contribution of bovine serum albumin (BSA) to the durability of the electrochemically synthesized poly(3,4-ethylenedioxythiophene) (PEDOT) film on a platinum (Pt) electrode. The electrode was capable to effectively adsorb the nano Au particles (AuNPs) to form a uniform layout, which was then able to immobilize the horseradish peroxidase (HRP) to construct a functional HRP/AuNPs/PEDOT(BSA)/Pt biosensor. Cyclic voltammetry was employed to evaluate the performance of the biosensor through the measurement of hydrogen peroxide. Our results revealed a satisfied linear correlation between the cathodic current and the concentration of H2O2. Furthermore, the addition of oxidized form of nicotinamide adenine dinucleotide, or NAD⁺, as the electron transfer mediator in the detection solution could dramatically enhance the sensitivity of detection by about 35.5%. The main advantages of the current biosensor are its durability, sensitivity, reliability, and biocompatibility.