These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Comparative analytic occlusal study of clinical use of T-scan systems]. Author: Patyk A, Lotzmann U, Scherer C, Kobes LW. Journal: ZWR; 1989 Sep; 98(9):752-5. PubMed ID: 2700863. Abstract: Occlusal corrections are dependent on a precise registration. The T-Scan system has recently been presented to the market as an innovative, computer-aided device, capable of providing exact information regarding position, strength and frequency of occlusal contacts. The diagnostic validity of this technique was tested in a comparing study on 16 patients. The occlusal situation of these patients was analyzed by use of the T-Scan system and the markings compared with those obtained from UV-sensitive articulation foil. The results show that the diagnostic possibilities of the T-Scan system are very much limited. The use oft he T-Scan system as the only means of occlusal analysis cannot be recommended. The main reason for this is an often misleading reproduction of occlusal contacts, caused by the T-Scan-sensor-foil being too thick and too inflexible and therefore creating an uncontrollable shift of the mandibula when closing the bite. However, the T-Scan system seems to be a valuable tool for education of both students and patients, as the on-screen coulour display of occlusal mechanisms are really impressive.[Abstract] [Full Text] [Related] [New Search]