These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Facile fabrication of an electrochemical aptasensor based on magnetic electrode by using streptavidin modified magnetic beads for sensitive and specific detection of Hg(2.).
    Author: Wu D, Wang Y, Zhang Y, Ma H, Pang X, Hu L, Du B, Wei Q.
    Journal: Biosens Bioelectron; 2016 Aug 15; 82():9-13. PubMed ID: 27031185.
    Abstract:
    In this work, a novel electrochemical aptasensor was developed for sensitive and specific detection of Hg(2+) based on thymine-Hg(2+)-thymine (T-Hg(2+)-T) structure via application of thionine (Th) as indicator signal. For the fabrication of the aptasensor, streptavidin modified magnetic beads (Fe3O4-SA) was firmly immobilized onto the magnetic glassy carbon electrode (MGCE) benefited from its magnetic character. Then biotin labeled T-riched single stranded DNA (Bio-ssDNA) connected with Fe3O4-SA specifically and steadily because of the specific binding capacity between streptavidin and biotin. The stable structure of T-Hg(2+)-T formed in the present of Hg(2+) provided convenience for the intercalation of Th. The detection of Hg(2+) was achieved by recording the differential pulse voltammetry (DPV) signal of Th. Under optimal experimental conditions, the linear range of the fabricated electrochemical aptasensor was 1-200nmol/L, with a detection limit of 0.33nmol/L. Furthermore, the proposed aptasensor may find a potential application for the detection of Hg(2+) in real water sample analysis.
    [Abstract] [Full Text] [Related] [New Search]