These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: A side-hole catheter for catheterization of a difficult internal mammary artery.
    Author: Oh JS, Choi BG, Chun HJ, Lee HG.
    Journal: Acta Radiol; 2017 Mar; 58(3):307-310. PubMed ID: 27083204.
    Abstract:
    Background The internal mammary artery (IMA) can be a source of hemoptysis in patients with chronic lung disease. Intervention via the IMA can be a challenge due to anatomic variations and lead to excessive contrast use and radiation exposure. Purpose To evaluate safety and efficiency of a new side-hole catheter for the catheterization of the IMA in patients with hemoptysis. Material and Methods From January 2011 to August 2014, a total of 96 transarterial embolization procedures required exact evaluation of the IMA due to chronic lung disease involving the anterior thorax. In 17 cases (18%) of these 96 procedures, the conventional selective IMA angiography failed and instead a novel side-hole catheter as a modification of a cobra-type curved catheter was used. The side hole allowed passage of a micro-wire and catheter. Results Failed catheterizations were due to severe vascular tortuosity, acutely angulated subclavian artery, or abnormal takeoff of the IMA. The Cobra shaped catheter with the microcatheter through the side-hole catheter yielded a technical success rate of 100%. Longer time was required to catheterize the IMA with the Cobra shaped catheter than with the side-hole catheter (17 vs. 2 min, P < 0.05). There were no procedure-related complications. Conclusion Side-hole catheter technique is useful in patients whose internal mammary artery is difficult to access. Further design revisions are needed to improve the ease and speed of IMA catheterization and angiography.
    [Abstract] [Full Text] [Related] [New Search]