These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Design and fabrication of an in situ gamma radioactivity measurement system for marine environment and its calibration with Monte Carlo method.
    Author: Abdollahnejad H, Vosoughi N, Zare MR.
    Journal: Appl Radiat Isot; 2016 Aug; 114():87-91. PubMed ID: 27213808.
    Abstract:
    Simulation, design and fabrication of a sealing enclosure is carried out for a NaI(Tl) 2″×2″ detector, to be used as in situ gamma radioactivity measurement system in marine environment. Effect of sealing enclosure on performance of the system in laboratory and marine environment (distinct tank with 10m(3) volume) were studied using point sources. The marine volumetric efficiency for radiation with 1461keV energy (from (40)K) is measured with KCl volumetric liquid source diluted in distinct tank. The experimental and simulated efficiency values agreed well. Marine volumetric efficiency calibration curve is calculated for 60keV to 1461keV energy with Monte Carlo method. This curve indicates that efficiency increasing rapidly up to 140.5keV but then drops exponentially.
    [Abstract] [Full Text] [Related] [New Search]