These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Performance Evaluation of the Becton Dickinson FACSPresto™ Near-Patient CD4 Instrument in a Laboratory and Typical Field Clinic Setting in South Africa. Author: Coetzee LM, Moodley K, Glencross DK. Journal: PLoS One; 2016; 11(5):e0156266. PubMed ID: 27224025. Abstract: BACKGROUND: The BD-FACSPresto™ CD4 is a new, point-of-care (POC) instrument utilising finger-stick capillary blood sampling. This study evaluated its performance against predicate CD4 testing in South Africa. METHODS: Phase-I testing: HIV+ patient samples (n = 214) were analysed on the Presto™ under ideal laboratory conditions using venous blood. During Phase-II, 135 patients were capillary-bled for CD4 testing on FACSPresto™, performed according to manufacturer instruction. Comparative statistical analyses against predicate PLG/CD4 method and industry standards were done using GraphPad Prism 6. It included Bland-Altman with 95% limits of agreement (LOA) and percentage similarity with coefficient of variation (%CV) analyses for absolute CD4 count (cells/μl) and CD4 percentage of lymphocytes (CD4%). RESULTS: In Phase-I, 179/217 samples yielded reportable results with Presto™ using venous blood filled cartridges. Compared to predicate, a mean bias of 40.4±45.8 (LOA of -49.2 to 130.2) and %similarity (%CV) of 106.1%±7.75 (7.3%) was noted for CD4 absolute counts. In Phase-2 field study, 118/135 capillary-bled Presto™ samples resulted CD4 parameters. Compared to predicate, a mean bias of 50.2±92.8 (LOA of -131.7 to 232) with %similarity (%CV) 105%±10.8 (10.3%), and 2.87±2.7 (LOA of -8.2 to 2.5) with similarity of 94.7±6.5% (6.83%) noted for absolute CD4 and CD4% respectively. No significant clinical differences were indicated for either parameter using two sampling methods. CONCLUSION: The Presto™ produced remarkable precision to predicate methods, irrespective of venous or capillary blood sampling. A consistent, clinically insignificant over-estimation (5-7%) of counts against PLG/CD4 and equivalency to FACSCount™ was noted. Further field studies are awaited to confirm longer-term use.[Abstract] [Full Text] [Related] [New Search]