These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Comparison of nonisotopic and radioimmunoassay techniques for digoxin: a study based on external quality assurance measurements. Author: Wilson JF, Tsanaclis LM, Williams J, Tedstone JE, Richens A. Journal: Ther Drug Monit; 1989; 11(4):477-9. PubMed ID: 2741197. Abstract: A comparison of the precision and accuracy of three nonisotopic and of radioimmunoassay (RIA) techniques for digoxin was made using data from 141 samples of spiked human serum assayed between 1984 and 1987 by members of the Heathcontrol external quality assurance scheme. There were significant differences between techniques in the number of observations rejected as outliers greater than 3 SD from the sample mean, the percentages of rejected measurements being Abbott TDX 1.4, RIA 3.3, Ames TDA 4.2, and Syva EMIT 6.3. Ames TDA and Syva EMIT were shown to have a significantly lower precision than the other techniques in measurements of digoxin concentrations less than 1 nmol/L. A significant 4% negative bias was observed for Ames TDA measurements of digoxin concentrations between 1 and 2.6 nmol/L. A concentration-related bias was demonstrated for Abbott TDX measurements that varied from 14% overestimates at concentrations less than 1 nmol/L to 6.7% underestimates of digoxin at concentrations greater than 2.6 nmol/L.[Abstract] [Full Text] [Related] [New Search]