These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Batch-fabricated thin-film electrodes for stimulation of the central auditory system.
    Author: Anderson DJ, Najafi K, Tanghe SJ, Evans DA, Levy KL, Hetke JF, Xue XL, Zappia JJ, Wise KD.
    Journal: IEEE Trans Biomed Eng; 1989 Jul; 36(7):693-704. PubMed ID: 2744793.
    Abstract:
    Silicon micromachining and thin-film technology have been employed to fabricate iridium stimulating arrays which can be used to excite discrete volumes of the central nervous system. Silicon multichannel probes with thicknesses ranging from 1 to 40 microns and arbitrary two-dimensional shapes can be fabricated using a high-yield, circuit-compatible process. Iridium stimulating sites are shown to have similar characteristics to iridium wire electrodes. Accelerated pulse testing with over 8 million 100 microA biphasic current pulses on 8000 microns 2 sites has demonstrated the long-term stability of iridium and activated iridium sites. In vivo tests have been performed in the central auditory pathways to demonstrate neural activation using the devices. These tests show a selective activation both as a function of site separation and site size.
    [Abstract] [Full Text] [Related] [New Search]