These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Effectiveness of the Trima Accel cell separator in the double dose plateletpheresis. Author: Keklik M, Korkmaz S, Kalan U, Sarikoc M, Keklik E. Journal: Transfus Apher Sci; 2016 Oct; 55(2):240-242. PubMed ID: 27522105. Abstract: The use of apheresis equipment to collect blood components has rapidly increased in the past years. A variety of apheresis instruments are now available on the market for double dose plateletpheresis. We evaluated an apheresis instrument, the Trima Accel, with regard to platelet (PLT) yield, collection efficiency (CE), and collection rate (CR) in a retrospective, randomized study in 110 donors. The Trima Accel cell separator efficiently collected double dose platelets with median PLT yields of 3.7 × 1011, mean CE of 74.99 ± 14.40% and mean CR of 0.096 ± 0.012 × 1011/min.[Abstract] [Full Text] [Related] [New Search]