  • Title: Freeze-Drying Process Development and Scale-Up: Scale-Up of Edge Vial Versus Center Vial Heat Transfer Coefficients, Kv.
    Author: Pikal MJ, Bogner R, Mudhivarthi V, Sharma P, Sane P.
    Journal: J Pharm Sci; 2016 Nov; 105(11):3333-3343. PubMed ID: 27666376.
    Abstract:
    This report presents calculations of the difference between the vial heat transfer coefficient of the "edge vial" and the "center vial" at all scales. The only scale-up adjustment for center vials is for the contribution of radiation from the shelf upon which the vial sits by replacing the emissivity of the laboratory dryer shelf with the emissivity of the production dryer shelf. With edge vials, scales-up adjustments are more complex. While convection is not important, heat transfer from the wall to the bands (surrounding the vial array) by radiation and directly from the band to the vials by both radiation and conduction is important; this radiation heat transfer depends on the emissivity of the vial and the bands and is nearly independent of the emissivity of the dryer walls. Differences in wall temperatures do impact the edge vial effect and scale-up, and estimates for wall temperatures are needed for both laboratory and manufacturing dryers. Auto-loading systems (no bands) may give different edge vial heat transfer coefficients than when operating with bands. Satisfactory agreement between theoretical predictions and experimental values of the edge vial effect indicate that results calculated from the theory are of useful accuracy.
    [Abstract] [Full Text] [Related] [New Search]