These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Wearable impedance monitoring system for dialysis patients.
    Author: Bonnet S, Bourgerette A, Gharbi S, Rubeck C, Arkouche W, Massot B, McAdams E, Montalibet A, Jallon P.
    Journal: Annu Int Conf IEEE Eng Med Biol Soc; 2016 Aug; 2016():5196-5199. PubMed ID: 28269435.
    Abstract:
    This paper describes the development and the validation of a prototype wearable miniaturized impedance monitoring system for remote monitoring in home-based dialysis patients. This device is intended to assess the hydration status of dialysis patients using calf impedance measurements. The system is based on the low-power AD8302 component. The impedance calibration procedure is described together with the Cole parameter estimation and the hydric volume estimation. Results are given on a test cell to validate the design and on preliminary calf measurements showing Cole parameter variations during hemodialysis.
    [Abstract] [Full Text] [Related] [New Search]