These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A Prospective Comparison of Transcutaneous and Serum Bilirubin Within Brief Time Intervals. Author: Jones DF, McRea AR, Knowles JD, Lin FC, Burnette E, Reller LA, Lohr JA. Journal: Clin Pediatr (Phila); 2017 Oct; 56(11):1013-1017. PubMed ID: 28366015. Abstract: The American Academy of Pediatrics recommends screening newborns ≥35 weeks' gestation with total serum bilirubin (TSB) or transcutaneous bilirubin (TcB) to detect hyperbilirubinemia. Retrospective studies show TcB measurements strongly correlate with TSB; however, few prospective trials document this relationship. Furthermore, Dräger's newest TcB instrument, JM-105, remains unstudied in the United States. We measure TcB on foreheads and sternums of newborns using JM-105 and Bilichek devices within 30 minutes of TSB measurement. We find best overall TcB/TSB correlation with JM-105 on the sternum (mean TcB-TSB difference: -0.21 ± 1.15 mg/dL). Correlations between paired measurements for TcB on the sternum using JM-105 were 0.93 for all TSB levels (n = 178), 0.82 for TSB > 10 (n = 19), 0.69 for TSB > 12 (n = 11), and 0.52 for TSB > 15 (n = 6). TcB accuracy via JM-105 on the sternum significantly differed among races ( P < .001). For 5% of paired measurements, TcB with JM-105 on the sternum underestimated TSB by ≥2 mg/dL, and for <1% by ≥3 mg/dL.[Abstract] [Full Text] [Related] [New Search]