These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Evaluation of the new C-arm guiding system ClearGuide® in an orthopaedic and trauma operating theatre. Author: Müller MC, Frege S, Strauss AC, Gathen M, Windemuth M, Striepens EN. Journal: Int J Med Robot; 2017 Sep; 13(3):. PubMed ID: 28371198. Abstract: BACKGROUND: The objective was to evaluate whether the new intraoperative C-arm guiding system ClearGuide® (CG) reduces radiation exposure of the staff in an Orthopaedic and Trauma operation theatre. METHODS: Data of 95 patients CG was used were retrospectively compared using matched-pair analysis with controls without CG. Radiation dose (RD), fluoroscopic time (FT) and operation time (OT) were analysed in ten types of operative procedures. RESULTS: Use of CG led to a significant reduction (p ≤ 0.05) of the RD in intramedullary nailing and plate fixation of femoral shaft fractures as well as plating of tibia shaft fractures. Concerning FT, use of CG led to a significant reduction (p ≤ 0.05) while performing kyphoplasties and plate fixation of femoral shaft fractures. Regarding OT, no statistical significance was observed. CONCLUSIONS: CG as a simple, reproducible and intuitive communication tool for C-arm guidance reduces intraoperative staff radiation exposure especially while fixation of long bone fractures and in spine surgery.[Abstract] [Full Text] [Related] [New Search]