These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Carotid-Femoral Pulse Wave Velocity Assessed by Ultrasound: A Study with Echotracking Technology. Author: Collette M, Palombo C, Morizzo C, Sbragi S, Kozakova M, Leftheriotis G. Journal: Ultrasound Med Biol; 2017 Jun; 43(6):1187-1194. PubMed ID: 28395967. Abstract: Described here is a new method for determination of carotid-femoral pulse wave velocity (PWV) based on arterial diameter waveform recording by an ultrasound system. The study was carried out on 120 consecutive patients. Carotid-femoral PWV was determined using a tonometric technique (PWVpp, PulsePen, DiaTecne, Milan, Italy) and an echotracking ultrasound system (PWVet, E-Track, Aloka, Tokyo, Japan). The relationship between PWVpp and PWVet was evaluated by linear regression and Bland-Altman analysis. There was excellent agreement between PWVet and PWVpp (Pearson's r = 0.94, 95% confidence interval: 0.91-0.96, p < 0.0001; PWVet = 0.88 × PWVpp + 0.57). The Bland-Altman plot revealed an offset of -0.33 m/s with limits of agreement from -2.21 to 1.54 m/s. The coefficients of variation for within-subject repeatability between PWVet and PWVpp had were 5.79% and 8.47%, respectively, without significant differences in the Bland-Altman analysis. The results suggest that echotracking technology can provide a reliable estimate of aortic stiffness comparable to that of the tonometric techniques.[Abstract] [Full Text] [Related] [New Search]