These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Automated temperature control system for vagal cooling.
    Author: Cha EJ, Chow E, Vega DM, Yamashiro SM.
    Journal: J Appl Physiol (1985); 1988 Jul; 65(1):469-72. PubMed ID: 2841281.
    Abstract:
    An automated cooling system for vagal cold blockade was developed. A simple electronic circuit is described that enabled regulation of the steady-state nerve or circulant temperature to within +/- 0.1 degree C by alternating the cold circulant between low and high flows (10 and 350 ml/min, respectively). The 90% rise time ranged from 15 to 40 s depending on the desired steady state and the surrounding temperatures. The present apparatus can be conveniently and safely used, especially for differential vagal cold blockade.