These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Introduction of a Novel Smartphone-Coupled Blood Glucose Monitoring System. Author: Jendrike N, Baumstark A, Chen CH, Rittmeyer D, Haug C, Freckmann G. Journal: J Diabetes Sci Technol; 2017 Nov; 11(6):1231-1233. PubMed ID: 28459160. Abstract: The novel system for self-monitoring of blood glucose (SMBG) PixoTest couples SMBG to a smartphone and does not require a separate glucose meter. The integrated system includes all components necessary for a glucose measurement, and owing to a colorimetric measurement principle, a smartphone camera can capture color changes and a software app calculates the corresponding glucose value. In the presented study, the system was evaluated in terms of system accuracy as described in ISO 15197:2013. It was shown to fulfill system accuracy requirements with 97-99% of results from three different reagent system lots within the accuracy limits and 100% of results within zone A of the consensus error grid.[Abstract] [Full Text] [Related] [New Search]