  • Title: Low Cryptococcus Antigen Titers as Determined by Lateral Flow Assay Should Be Interpreted Cautiously in Patients without Prior Diagnosis of Cryptococcal Infection.
    Author: Dubbels M, Granger D, Theel ES.
    Journal: J Clin Microbiol; 2017 Aug; 55(8):2472-2479. PubMed ID: 28566315.
    Abstract:
    Detection of Cryptococcus antigen (CrAg) is invaluable for establishing cryptococcal disease. Multiple different methods for CrAg detection are available, including a lateral flow assay (LFA). Despite excellent performance of the CrAg LFA, we have observed multiple cases of low-titer (≤1:5) positive CrAg LFA results in patients for whom cryptococcosis was ultimately excluded. To investigate the accuracy of low-titer positive CrAg LFA results, we performed chart reviews for all patients with positive CrAg LFA results between June 2014 and December 2016. During this period, serum and/or cerebrospinal fluid (CSF) samples from 3,969 patients were tested with the CrAg LFA, and 55 patients (1.5%) tested positive. Thirty-eight of those patients lacked a history of cryptococcal disease and were the focus of this study. Fungal culture or histopathology confirmed Cryptococcus infection for 20 patients (52.6%), and CrAg LFA titers in serum and CSF samples ranged from 1:5 to ≥1:2,560. For the 18 patients (47.4%) without culture or histopathological confirmation, the CrAg LFA results were considered true-positive results for 5 patients (titer range, 1:10 to ≥1:2,560), due to clinical improvement with targeted therapy and decreasing CrAg LFA titers. The remaining 13 patients had CrAg LFA titers of 1:2 (n = 11) or 1:5 (n = 2) and were ultimately diagnosed with an alternative condition (n = 11) or began therapy for possible cryptococcosis without improvement (n = 2), leading to an overall CrAg LFA false-positive rate of 34%. We recommend careful clinical correlation prior to establishing a diagnosis of cryptococcal infection for patients with first-time positive CrAg LFA titers of 1:2.
    [Abstract] [Full Text] [Related] [New Search]