    Abstract:
    PURPOSE: We aimed to assess polyethylene liners of retrieved hips of one design of a dual mobility (DM) cup liner and two designs of femoral stems to better understand the role of femoral stem design on polyethylene impingement. METHODS: This was a case-control study involving 70 retrieved highly cross-linked polyethylene (X3) liners used with ABGII (n = 35) and Rejuvenate (n = 35) stems (Stryker). All polyethylene liners were assessed for evidence of rim deformation and the damage quantified using metrology methods. RESULTS: A total of 80% of polyethylene liners paired with ABGII necks had macroscopic evidence of neck impingement resulting in a raised lip whilst 23% of liners paired with Rejuvenate necks had evidence of a raised lip (p < 0.0001). The height of the raised rims of the DM cups paired with ABGII necks had a median (range) of 139 μm (72-255). The height of the raised rims of the DM cups paired with Rejuvenate necks had a median (range) of 52 μm (45-90) (p < 0.0001). CONCLUSION: Our new findings from retrieved dual mobility bearings showed that polyethylene liner rim deformation resulting from impingement with the femoral neck occurs in early in-human function, is circumferential in distribution, and is affected by the stem neck design. We recommend the use of highly polished and non-edged neck designs when used in conjunction with DM cups.
    [Abstract] [Full Text] [Related] [New Search]