These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: On the Nature of Mass Spectrometer Analyzer Contamination.
    Author: Kang Y, Schneider BB, Covey TR.
    Journal: J Am Soc Mass Spectrom; 2017 Nov; 28(11):2384-2392. PubMed ID: 28733968.
    Abstract:
    Sample throughput in electrospray ionization mass spectrometry (ESI-MS) is limited by the need for frequent ion path cleaning to remove accumulated debris that can lead to charging and general performance degradation. Contamination of ion optics within the vacuum system is particularly problematic as routine cleaning requires additional time for cycling the vacuum pumps. Differential mobility spectrometry (DMS) can select targeted ion species for transmission, thereby reducing the total number of charged particles entering the vacuum system. In this work, we characterize the nature of instrument contamination, describe efforts to improve mass spectrometer robustness by applying DMS prefiltering to reduce contamination of the vacuum ion optics, and demonstrate the capability of DMS to extend the interval between mass spectrometer cleaning. In addition, we introduce a new approach to effectively detect large charged particles formed during the electrospray ionization (ESI) process. Graphical Abstract ᅟ.
    [Abstract] [Full Text] [Related] [New Search]