These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: YSZ-based sensor using Cr-Fe-based spinel-oxide electrodes for selective detection of CO. Author: Anggraini SA, Fujio Y, Ikeda H, Miura N. Journal: Anal Chim Acta; 2017 Aug 22; 982():176-184. PubMed ID: 28734357. Abstract: A selective carbon monoxide (CO) sensor was developed by the use of both of CuCrFeO4 and CoCrFeO4 as the sensing electrode (SE) for yttria-stabilized zirconia (YSZ)-based potentiometric sensor. The sensing-characteristic examinations of the YSZ-based sensors using each of spinel oxides as the single-SE sensor showed that CuCrFeO4-SE had the ability to detect CO, hydrocarbons and NOx gases, while CoCrFeO4-SE was sensitive to hydrocarbons and NOx gases. Thus, when both SEs were paired as a combined-SEs sensor, the resulting sensor could generate a selective response to CO at 450 °C under humid conditions. The sensor was also capable of detecting CO in the concentration range of 20-700 ppm. Its sensing mechanism that was examined via polarization-curve measurements was confirmed to be based on mixed-potential model. The CO response generated by the combined-SEs sensor was unaffected by the change of water vapor concentration in the range of 1.3-11.5 vol% H2O. Additionally, the sensing performance was stable during 13 days tested.[Abstract] [Full Text] [Related] [New Search]